Contents
Chapter 1 Product structure Chapter 2 Accessories list Chapter 3 Product informaion Chapter 4 Installation
Chapter 5 Operation
Chapter 6 Machine maintenance
Laser danger
CE Label
1. CO2 main unit | 2. Guide light arm | 3. Power cord |
4. Key | 5. Foot switch | 6. Fractional laser head |
7. Vaginal tightening head | 8. pulse head | 9. funnel |
1.Co2 laser whitening skin rejuvenation principle
1 PRINCIPLE OF СО2
1 BEFORE fi AFTER
Tl Cl Т2 С2
Da,yl
DayS
DaylO
Da,y15
в.еfо,ге After Веfоге Afteг
Co2 laser advantage
More user-friendly: Due to the controllability of the CO2 laser, the size and diameter of the treatment hole can be modified according to the different conditions of the individual skin, and the best for you is the best.
Safe and effective: the CO2 laser is certified by the national safety and is treated by a professional doctor. The experienced doctor can bring you the safest and most effective treatment according to the previous clinical technology.
Small healing of the wound: Since the beam of the laser is very small, the wound is easy to heal, and the uniform application makes the spot not overlap, so the skin can be quickly repaired when the whole part is applied.
Short recovery period: Compared with the earlier beauty project, the recovery period of the laser is shorter, it takes only 3-6 days to complete the process of molting, the appearance of the epidermis is naturally repaired, and the collagen growth under the skin lasts for 6 months. , no pain, does not affect work and study
3, special reminder after surgery:
Lattice laser treatment of wounds can generally heal within 8-24 hours, most patients do not need special medical care, with the repair of growth factors to promote rapid repair, just avoid contact with water, prevent infection before dislocation. ☆ 100% to avoid sun exposure ☆ sunscreen work is done to reduce the chance of pigmentation after inflammation, darker skin should pay attention to avoid sun exposure.
The difference between CO2 laser and pixel laser
Vaginal tightening laser is the use of water uptake 10600nm gold standard laser, using a three-dimensional lattice technology and 360-degree circular emission perfect combination of technology, to produce 50-70 degrees Celsius in a controlled depth of the vaginal mucosa, stimulate lamina propria and myometrium newborn fibroblasts, and damaged collagen fibers, elastic fibers, such as restructuring, so as to achieve vaginal wall thickening and tighteningof the vagina so that firmness, sensitivity enhancement, vagina tight as ever.
Vaginal tightening laser does not produce skin burns due to high temperature, each capable of Homogeneous heat, acting on a few hundred nanometers thickness Mucosa organization.
thermal radiation make Mucosa tissue contraction and collagen fibers thereby improving and restructuring vaginal tightening.
Shrink vagina: rapid tightening, constriction, lasting firmness.
Beautify vagina: dilute pigment, pink labia.
Moist vagina: increase secretion, eliminate dryness.
Maintenance vagina: deep rejuvenation, prevent aging.
To improve sensitivity.
Private health, balance PH, improve the internal environment.
Wavelength: | 10600nm |
Laser apparatus: | Sealed off laser device simulated by direct current |
Optical beam quality: | TEM00 |
Output power: | 60w |
Spot size: | 0.02~0.05mm2 |
Aiming beam: | Red semiconductor laser (635nm, less than 5mw) |
Beam transport device: | Articulated arm with 7-joint and balance weight |
Working mode: | Continuous operation |
Output Mode: | Continuous, single pulse, interval pulse and super pulse |
Scan scope: | Max 20mmx20mm |
Scan graphics: | Circle, triangle, square, rectangle,hexagon, ellipse, line |
Scanning Mode: | Random , Normal and Midsplit scan |
Scanning speed: | More than 10m/s |
Energy: | 1- 60W |
Distance: | 0.1-2.6mm | |
Cooling system: | air circulation | |
Weight: | 60kg | |
Dimension: | 65*48*117cm | |
Power: | 220V/11V |
1. Every parts name.
1
2
3
4
5
6
joints of seven light guide arm
By seven joint light guide arm to complete the laser beam transmission.
The light guide arm is the role of the laser transmission to patients to operation site
(target).
scanner
Graphics scanner can scan output, rectangular, round, hexagonal, triangular and other graphics, the largest area of about 20 * 20mm2 scanning;
Main unit
It includes a high voltage power supply, a low-voltage power supply, control system, cooling system, laser system and blowing system.
touch display panel
Mainly used for parameter setting
key switch
A key into the key switch, go to the "open" position, the instrument power on, go to the "off” position to cut the power, turn off the instrument.
10
11
9
8
7
wheels
In Jection hole
Overflow hole
Drainage hole
The power input socket
Foot switch
2. Installation
Checking
Take out the packing list after opening the box, then take out the host and spare parts according to the packing list, check every spare parts, wire connections, confirmed no abnormal or troubleshooting before installation, if have any problem, please contact manufacturer timely, so that resolve the problem Quickly.
Fill distilled water (if you keep the machine in cold place which is less than 5℃,advice anti -42℃ fluid install distilled water )
Press the stainless steel sheet at the above of the overflow hole, pull the water tube into the hole.
Put the tube into the In Jection hole, do like as above said, fill distilled water, it's ok when
the water overflow the Overflow hole; after turn on the machine for 3 minutes, the machine need fill water again.
Attention: you can't pull out the tube from the In Jection hole when using the machine.
You
need to change the water in every 60 days.
Installation of guide light arm
Guide light system was made up of guide light arm and connection seat .connection seat has fixed into the top of the shell. After check the guide light arm, connect it to connection seat, then fix the screws firmly.
Attention: guide light arm only can move to the arrow direction.
Move B 180 degree according to C, then press A according to the Arrow. Then insert B to the hole on main unit, then screw D, fix the guide light arm firmly.
Installation of the scan head
Whirl the scan head to clockwise direction.
Installation of control wire on the scan head, insert the control wire, then fix the screws firmly.
Installation of power cord and foot switch
Insert the power cord to the connector on the back of the machine, then check if firmly.
Connect the foot switch, aim at the gap, whirl to the clockwise direction, check if firmly.
change water
Put “add water” and ”run off water” connect, ventage plug connect run off water, then add water again.
Turn on the power of device.
Appear to be the stand-by screen as below when key switch is ON.
Select a mode (Fractional/Normal) and range of adjustment when the stand-by
screen comes up.
Operate it by stepping FOOT SWITCH after completing every procedure
Detail instructions of each mode are as follows.
Fractional mode: Light / Mid / Deep/Fusion )
Light
| |||
Mid
| |||
Deep
| |||
Fusion
| |||
| fractional resurfacing | ||
Superficial texture variations | |||
Fine lines and wrinkles |
1) Regulate energy: the background color will become purple, then press”+” or “-” at the right side, it will increase or decrease. * Range of Adjustment: 1-60W. | |
2)Adjust out point speed: press “+” or “-” to adjust. * Range of Adjustment: 0.1-10.0ms. Advice 1 | |
3) Adjust the graphic size. press “+” or”-” to adjust. * Range of 0.1-20mm | |
4) press”+” or “-” to adjust. Range of Adjustment: 0.1mm ~ 2.6mm. Advice 1.0 | |
6)Laser dwell time. press “+” or “-” to adjust. * Range of Adjustment: 1-20ms. Advice 1.0 | |
Gernerally choose “random” | 5)Adjust out point mode. Norbmal mode: scan point from side to side in turn Random modes: scanning point no order, randomly Midsplit modes: the next scan point scanning point and peers to keep half the distance matrix were. |
7)You could adjust the pattern shape here. pres s these four button, you can change different Shape,like |
Square, round, triangular, hexagonal, straight. | |
8) you can change the pattern shape size when press “+”,”-”, And preview if the size and shape is suitable when press “preview”. | |
9) After shooting laser beam, turn to Main screen when touch BACK box at the bottom of screen. | |
10)Touch Laser shot times, generally choose 1th | |
| 11)Stop and Start |
1. ACNE SCAR / SCAR REMOVE ( APPLY ANESTHETIC ON TREATMENT AREA FOR 30 MINS BEFORE TREATMENT ) 30 DAYS A TIME, 3-12 TIMES.
| |||
RING(MJ) | 8-12W | ||
REPEAT (SECONDS) | 1.0MS | ||
CORE(MJ) | 1.0MS | ||
RING SIZE | 0.6MM | ||
SCAN MODES | RANDOM | ||
REPEAT | 1TH |
2. SKIN REJUVENATION 30 DAYS A TIME, 3-6 TIMES.
| |||
RING(MJ | 3-5W | ||
REPEAT (SECONDS) | 1.0MS | ||
CORE(MJ) | 1.0MS | ||
RING SIZE | 0.8MM | ||
SCAN MODES | RANDOM | ||
REPEAT | 1TH | ||
3. ACNE REMOVE 30 DAYS A TIME, 1-3 TIMES.
| |||
RING(MJ | 5-9W | ||
REPEAT (SECONDS) | 1.0MS | ||
CORE(MJ) | 1.0MS | ||
RING SIZE | 0.8MM | ||
SCAN MODES | RANDOM | ||
REPEAT | 1TH |
Could not Use “Surgical” mode to operate these handle, because “Surgical” light will damage these handle glasses and head set program, it will be about 1000usd.
Instruction of cutting ( Surgical mode)
RETURN
Laser interval time
Laser dwell time
Laser energy
LASER OUTPUT MODE
choose “power”,then press
,
, 1-30W, you can adjust different power.
,
choose “pulse width”, then press bottom of the screen) will change.
, 1-10ms, the pulse frequency(at the
choose “interval time”, then press“ ”、“ ”, 1-10ms, the pulse frequency
will change.
When choose “pulse” mode, decrease “pulse width” or “interval time”, the frequency is more than 100hz/s, ultra pulse will change to “ultra pulse” mode automatically.
when choose “ultra pulse”, increase “decrease “pulse width” or “interval time”, the frequency is less than 100hz/s, ultra pulse will change to “pulse” mode automatically.
when choose “single” mode, adjust suitable power, and pulse width, press the foot switch, the Head will scan one point.
when choose “CW” mode, adjust power and pulse width, you can scan a straight line.
Could not Use “Sugical” mode to operate these handle, because “Sugical” light will damage these handle glasses and head set program, it will be about 1000usd.
Instruction of vaginal tightening ( Private mode ).
Touch “Power “ box, the background color will become purple, then press”+” or “-” on the right side, it will increase or decrease.
Range of Adjustment: 1-30W.
Touch “ Duration” box, press “+” or “-” to adjust.
Range of Adjustment: 0.1-10.0ms
Touch “Interval” box, press “+” or”-” to adjust.
Range of Adjustment: 1-5000ms
Touch “distance” box, press”+” or “-” to adjust.
Range of Adjustment: 0.1mm ~ 2.6mm
Touch “Time” box, press “+” or “-” to adjust.
Range of Adjustment: 1-20th.
8)
you can change the shape size when press “+”,”-”, And preview if the size and shape is suitable when press “preview”.
9)
pres s this two button, you can change different Shape,like Square, round, triangular, hexagonal, straight.
After shooting laser beam, turn to Main screen when touch BACK box at the bottom Of screen.
This parameter is for inside the vaginal.
Insert the support to the vagina, then insert the 360 degree head to the bottom of the support, pay attention to the number on the 360 degree head, press the foot switch, then pull out 0.5cm, press the foot switch again, then pull our 0.5cm again, press the foot switch again, do again, after the 360 degree head reach vulva, insert to the bottom of the support again, and do like the first time, total 3 times.
VAGINAL TIGHTENING
60-120 DAYS A TIME, 1-3 TIMES.
Before After
POWER
15-20W
DURATION
1.0MS
POINT NUMBER
7
DISTANCE
1.5MM
SCAN MODES
RANDOM
SCAN TIMES
1TH
SCAN ROWS
7
Preparation:
Alcohol, cotton ball, sterile treatment head.
Operation:
.
'.:·:.. .....-. .... .z '•_..., • ,
..
.
tightening and nipple colour lightening ( Vulvae mode).
Nipple colour lightening 30-60 days a time, 3 times.
Before After | |||
POWER | 7-9W | ||
DURATION | 1.0MS | ||
POINT NUMBER | 7 | ||
DISTANCE | 1.5MM | ||
SCAN MODES | RANDOM | ||
SCAN TIMES | 1TH | ||
SCAN ROWS | 7 |
LABIA MAGORA & LABIA MINORA COLOUR LIGHTING AND TIGHTENING 30-60 DAYS A TIME, 3 TIMES.
Before After | |||
POWER | 7-9W | ||
DURATION | 1.0MS | ||
POINT NUMBER | 7 | ||
DISTANCE | 1.5MM | ||
SCAN MODES | RANDOM | ||
SCAN TIMES | 1TH | ||
SCAN ROWS | 7 |
Operation attention:
1. Please fill anti-icing fluid instead of water when Winter is coming. ( anti-icing fluid is necessary when room temperature is under 5℃: |
2. Please test machine before using it. Advice to use machine for more than 5 minutes every month, otherwise the machine will lose activity and energy will be attenuated. |
3. The signal cable cannot be unplugged or install again while the machine power is on! Otherwise, probe and board will be burnt together! |
4. Glass lenses cannot be cleaned with alcohol or high temperature disinfection; only dry cotton swabs or a small amount of clean water can be used. |
5. The power cable and socket should be connected with ground wire, otherwise reduce life time of the probe. |
6. Please keep the glass away from dust. |
7. The Surgical MODE could not use Fractional probe. Otherwise, the glasses of Fractional probe will get burnt because of Surgical MODE is very high energy. And the Fractional probe will cost 1200usd a set. |
8. This probe only match this machine, installing to other machine will burn this probe. And this head cost 1200usd a set. |
9. Please install Fractional head after uninstall the cutting probe completely, Otherwise, Fractional head energy would be very small. |
Maintenance
The instrument's power supply and laser tube generate high voltage. Ask the professional to take care of the instrument to avoid electric shock.
lens cleaning
After six months of use, the output energy of the instrument will drop slightly, which may be caused by contamination of the focus lens of the cutter head. Wipe the lens several times with a damp cotton ball and carefully damage the lens.
chassis cleaning
If there is dust inside the chassis, use a damp cotton cloth to clean the cleaner or toothpaste. The cloth should not be too wet to prevent the instrument from shorting.
power calibration
The laser output power should be calibrated every year. A standard laser power meter should be used for calibration. The calibration personnel should have corresponding qualifications.
fuse replacement
Turn off the power and unplug the power cord before replacing the fuse. When replacing, use a screwdriver to unscrew the fuse mount, remove the used fuse, install a new fuse, install the fuse mount and tighten, note that the corresponding fuse specifications should be consistent with the nameplate.